Contact-Handled Bay Waste Related Volatile Organics
Monitoring History
On January 1, 2007, the WIPP initiated
monitoring for volatile organic compounds in the air around
contact-handled transuranic waste shipping containers as they are
unloaded for disposal. These results will be updated quarterly.
|
Analytes of Interest
|
Carbon tetrachloride and
1,1,1-trichloroethane |
|
Current instruments in use
|
Photovac
Voyager field Gas Chromatographs with electron capture detectors
|
|
Limits of detection (Minimum
Per MFG)
|
better
than 0.05 ppm (vol) for both compounds |
|
Sample count for the quarter
ending 12-31/2011
|
7519
samples from 10/1/2011 through 12/31/2011 |
|
Count of samples with VOC's detected
(excluding calibration runs)
|
CC14: 14
1,1,1-TCA: 0 |
|
Average concentration of peaks
detected (peaks quantifiable)
|
CC14:
0.034
1,1,1-TCA: n/d
|
|